If the Mach-1 Motion application fails to load and displays error -2633: "LabVIEW: Attempted to get Child Nodes on invalid node refnum", this may be due to a missing multiaxis load cell calibration file (FTXXXXX.cal). Solution 1: Reset Multiaxis Load ...
There have been a few issues noticed with specifically Trend Micro antivirus software: https://www.trendmicro.com/. In particular, this antivirus software interferes with incoming force and position data from the Mach-1. Some examples include: ...
Here are the main differences when using a spherical vs cylindrical indenter: Spherical Shaped Indenter Cylindrical Shaped Indenter A model should be used to fit to this data (Hayes, Hertz, Sneddon). See Mach-1 Analysis User Manual for more ...
Note that for version 4.5.0.1 of Mach-1 Motion or later with an ESP302 controller + interconnection board + load cell smart connectors, the load cell calibration matrix is hardcoded and must be modified by a Biomomentum technician. How are multi-axis ...
The general Error -88705 can be solved using the following method: https://support.biomomentum.com/portal/en/kb/articles/daqmx-self-test-error-88705 In this specific case of the error, there is a blank screen (no load cell data is shown) for a few ...
NOTE: in Mach-1 Motion version 4.5.0.1+ (ESP302), the software will automatically search through all DAQ ports. The method below is not necessary. This error is usually caused by having an incorrect DAQ# in the Mach-1 Motion configuration file. ...
To view Biomomentum's Knowledge Base articles visible only to registered users, you must create an account. Click "Sign Up" in the top right of the screen to get started.
The general Error -88705 can be solved using the following method: https://support.biomomentum.com/portal/en/kb/articles/daqmx-self-test-error-88705 In this specific case of the error, there is a blank screen (no load cell data is shown) for a few ...
In the case where the stage does not move when "Up" or "Down" are clicked, it is likely that the stage has been turned OFF. To turn the stage back ON, click "Enable Axis" or "Enable Axes" in the "Manual Controls" window. Stages will automatically ...
Error code: -1073807339 can appear for multiple reason. The most common reason is that the ESP controller power button is off. To fix this error, reach behind the Motion Controller and activate the switch powering ON the ESP controller. Other ...