Verify the actuator controller connection and/or power (Error code: -1073807339)

Verify the actuator controller connection and/or power (Error code: -1073807339)

Error code: -1073807339 can appear for multiple reason. The most common reason is that the ESP controller power button is off.
To fix this error, reach behind the Motion Controller and activate the switch powering ON the ESP controller.


Other possible reasons:
-The controller has not been power cycled since the PC was restarted. The solution in this case is to turn the controller OFF and ON again.
-On a new installation or after reinstallation of the system, it is possible that the COM# has changed. This COM# can be modified in the Mach-1 Motion Configuration/Hardware menu when opening Mach-1 Motion. Valid numbers are usually between 1 and 6.
-There is also the possibility that the "Cable to DAQ Card" is loose and must be tightened.


Problem associated with updating from Windows 10 to Windows 11:
-An issue was discovered with the COM port driver when updating from Windows 10 to Windows 11.
-To solve this particular error
  1. open "Device Manager" on the PC.
  2. Use the drop down arrow to the left of the "Ports (COM & LPT) line to show the COM ports.
  3. Right-click the COM port the system is using (in older systems this is usually COM1).
  4. Click "Update driver" > "Search automatically for drivers" > "close".
  5. Restart the PC.
  6. Test the connection by opening the Mach-1 Motion application and seeing if live data is shown.

    • Related Articles

    • Basler Pylon Gain Out Of Range (Error code: 1172)

      NOTE: SOLVED IN VERSION 4.4.0.29 The solution to this error is to reduce the gain of the camera. SOLUTION: 1) Open Mach-1 Motion 2) Open camera (error occurs) 3) Open camera settings window 4) Reduce gain to <12. 5) Restart Mach-1 Motion [in one case ...
    • DAQmx Self-Test (Error: -88705)

      This error is usually caused by having an incorrect DAQ# in the Mach-1 Motion configuration file. Windows can sometimes change this number during a Windows Update. To fix this, the DAQ number can be changed in the Mach-1 Motion Configuration/Hardware ...
    • Stage serial number mismatch (Error: 5002)

      Each Mach-1 system is configured with a specific XML configuration file for each configuration. This error is caused by trying to open the incorrect combination of stages plugged in to the ESP controller and/or opening the incorrect configuration ...
    • Camera related errors (Error: 5003, Error: -200279)

      The camera integration into Mach-1 Motion is very intensive on the PC. If Mach-1 Motion is not able to run smoothly and starts losing data, an error will appear. To minimize the likelihood of camera errors, try the following suggestions: ALWAYS save ...
    • C Drive Disk full (Error code: 9 and Error code: -1073807202)

      If the Mach-1 Motion application is not properly loading, it may be because your C: drive has run out of space. The common errors associated with this are -1073807202 "LabVIEW: (Hex 0xBFFF009E)" and 9 "LabVIEW: Disk full.". To solve this, follow the ...