DAQmx Self-Test (Error -88705)

DAQmx Self-Test (Error: -88705)

This error is usually caused by having an incorrect DAQ# in the Mach-1 Motion configuration file. Windows can sometimes change this number during a Windows Update.
To fix this, the DAQ number can be changed in the Mach-1 Motion Configuration/Hardware menu. Valid numbers usually fall between 1 to 6.
Please change the number, open Mach-1 Motion and verify there is no error. Repeat until Mach-1 Motion initializes normally.


If there are still issues after trying different DAQ numbers, please restart the PC in case the issue was regarding the DAQ PCIe card driver instead.
    • Related Articles

    • Camera related errors (Error: 5003, Error: -200279)

      The camera integration into Mach-1 Motion is very intensive on the PC. If Mach-1 Motion is not able to run smoothly and starts losing data, an error will appear. To minimize the likelihood of camera errors, try the following suggestions: ALWAYS save ...
    • Verify the actuator controller connection and/or power (Error code: -1073807339)

      Error code: -1073807339 can appear for multiple reason. The most common reason is that the ESP controller power button is off. To fix this error, reach behind the Motion Controller and activate the switch powering ON the ESP controller. Other ...
    • Stage serial number mismatch (Error: 5002)

      Each Mach-1 system is configured with a specific XML configuration file for each configuration. This error is caused by trying to open the incorrect combination of stages plugged in to the ESP controller and/or opening the incorrect configuration ...
    • Basler Pylon Gain Out Of Range (Error code: 1172)

      NOTE: SOLVED IN VERSION 4.4.0.29 The solution to this error is to reduce the gain of the camera. SOLUTION: 1) Open Mach-1 Motion 2) Open camera (error occurs) 3) Open camera settings window 4) Reduce gain to <12. 5) Restart Mach-1 Motion [in one case ...
    • No data stream when opening Mach-1 Motion, then Error -88705 after a few minutes

      The general Error -88705 can be solved using the following method: https://support.biomomentum.com/portal/en/kb/articles/daqmx-self-test-error-88705 In this specific case of the error, there is a blank screen (no load cell data is shown) for a few ...