DAQmx Self-Test (Error -88705)

DAQmx Self-Test (Error: -88705)

This error is usually caused by having an incorrect DAQ# in the Mach-1 Motion configuration file. Windows can sometimes change this number during a Windows Update.
To fix this, the DAQ number can be changed in the Mach-1 Motion Configuration/Hardware menu. Valid numbers usually fall between 1 to 6.
Please change the number, open Mach-1 Motion and verify there is no error. Repeat until Mach-1 Motion initializes normally.


If there are still issues after trying different DAQ numbers, please restart the PC in case the issue was regarding the DAQ PCIe card driver instead.
    • Related Articles

    • Verify the actuator controller connection and/or power (Error code: -1073807339)

      Error code: -1073807339 can appear for multiple reason. The most common reason is that the ESP controller power button is off. To fix this error, reach behind the Motion Controller and activate the switch powering ON the ESP controller. Other ...
    • Stage serial number mismatch (Error: 5002)

      Each Mach-1 system is configured with a specific XML configuration file for each configuration. This error is caused by trying to open the incorrect combination of stages plugged in to the ESP controller and/or opening the incorrect configuration ...
    • Basler Pylon Gain Out Of Range (Error code: 1172)

      NOTE: SOLVED IN VERSION 4.4.0.29 The solution to this error is to reduce the gain of the camera. SOLUTION: 1) Open Mach-1 Motion 2) Open camera (error occurs) 3) Open camera settings window 4) Reduce gain to <12. 5) Restart Mach-1 Motion [in one case ...
    • C Drive Disk full (Error code: 9 and Error code: -1073807202)

      If the Mach-1 Motion application is not properly loading, it may be because your C: drive has run out of space. The common errors associated with this are -1073807202 "LabVIEW: (Hex 0xBFFF009E)" and 9 "LabVIEW: Disk full.". To solve this, follow the ...
    • Low-Pass Filter Questions

      NOTE: THIS HAS BEEN SOLVED IN VERSION 4.4.0.29 BY APPLYING THE FILTER TO BOTH LOAD AND POSITION CHANNELS. In the "Load Cells" menu, the Low-Pass Filter can be activated or deactivated. Some common values are 3rd order 5Hz cutoff and 3rd order 20Hz ...